* File Name: HeartRateProcessing.c
|
|
*
|
|
* Version: 1.0
|
|
*
|
|
* Description:
|
|
* This file implements the heart rate measurement capability in the the PSoC 4
|
|
* BLE Lab 3.
|
|
*
|
|
* Hardware Dependency:
|
|
* CY8CKIT-042 BLE Pioneer Kit
|
|
*
|
|
******************************************************************************
|
|
* Copyright (2014), Cypress Semiconductor Corporation.
|
|
******************************************************************************
|
|
* This software is owned by Cypress Semiconductor Corporation (Cypress) and is
|
|
* protected by and subject to worldwide patent protection (United States and
|
|
* foreign), United States copyright laws and international treaty provisions.
|
|
* Cypress hereby grants to licensee a personal, non-exclusive, non-transferable
|
|
* license to copy, use, modify, create derivative works of, and compile the
|
|
* Cypress Source Code and derivative works for the sole purpose of creating
|
|
* custom software in support of licensee product to be used only in conjunction
|
|
* with a Cypress integrated circuit as specified in the applicable agreement.
|
|
* Any reproduction, modification, translation, compilation, or representation of
|
|
* this software except as specified above is prohibited without the express
|
|
* written permission of Cypress.
|
|
*
|
|
* Disclaimer: CYPRESS MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH
|
|
* REGARD TO THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED
|
|
* WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
|
|
* Cypress reserves the right to make changes without further notice to the
|
|
* materials described herein. Cypress does not assume any liability arising out
|
|
* of the application or use of any product or circuit described herein. Cypress
|
|
* does not authorize its products for use as critical components in life-support
|
|
* systems where a malfunction or failure may reasonably be expected to result in
|
|
* significant injury to the user. The inclusion of Cypress' product in a life-
|
|
* support systems application implies that the manufacturer assumes all risk of
|
|
* such use and in doing so indemnifies Cypress against all charges. Use may be
|
|
* limited by and subject to the applicable Cypress software license agreement.
|
|
*****************************************************************************/
|
|
|
|
|
|
/*****************************************************************************
|
|
* Included headers
|
|
*****************************************************************************/
|
|
#include <project.h>
|
|
#include <stdbool.h>
|
|
#include "WatchdogTimer.h"
|
|
|
|
|
|
/*****************************************************************************
|
|
* Macros
|
|
*****************************************************************************/
|
|
#define HEART_RATE_CHANNEL (0)
|
|
#define ADC_THRESHOLD (0x06A0)
|
|
#define SEC_IN_MIN (60)
|
|
#define MS_TO_SECOND (1000)
|
|
|
|
|
|
/*****************************************************************************
|
|
* Public variables
|
|
*****************************************************************************/
|
|
uint8 heartRate = 0;
|
|
|
|
|
|
/*****************************************************************************
|
|
* Public function definitions
|
|
*****************************************************************************/
|
|
|
|
/*****************************************************************************
|
|
* Function Name: ProcessHeartRateSignal
|
|
******************************************************************************
|
|
* Summary:
|
|
* Measures the heart rate of the user.
|
|
*
|
|
* Parameters:
|
|
* None
|
|
*
|
|
* Return:
|
|
* None
|
|
*
|
|
* Theory:
|
|
* This function acts as a simple model for heart rate measurement.
|
|
* The function takes the ADC sampled output and compares it to a threshold.
|
|
* If the ADC output is more than the threshold then it is considered as a
|
|
* valid beat (R peak). The rising edge of this R peak is identified and the
|
|
* corresponding system timestamp is noted. The RR-interval between two
|
|
* peaks is then calculated and converted to a heart rate value in beats per
|
|
* minute. The RR-interval period is calculated over a rolling window.
|
|
*
|
|
* Side Effects:
|
|
* None
|
|
*
|
|
*****************************************************************************/
|
|
void ProcessHeartRateSignal(void)
|
|
{
|
|
static bool newBeat = true;
|
|
static bool firstTime = true;
|
|
static uint32 previousBeatTime = 0;
|
|
static uint32 newBeatTime = 0;
|
|
int16 adcOut;
|
|
uint32 twoSampleTime = 0;
|
|
|
|
/* Get the ADC output */
|
|
ADC_StartConvert();
|
|
ADC_IsEndConversion(ADC_WAIT_FOR_RESULT);
|
|
adcOut = ADC_GetResult16(HEART_RATE_CHANNEL);
|
|
|
|
/* If the ADC output is more than a fixed threshold, consider that a
|
|
* valid R peak */
|
|
if (adcOut > ADC_THRESHOLD)
|
|
{
|
|
/* Check if the R peak just started - i.e. identify the rising
|
|
* edge of the R peak */
|
|
if(newBeat)
|
|
{
|
|
/* Check if this is the first R-peak seen by the device yet.
|
|
* If that is the case, we cannot calculate a heart rate value
|
|
* yet since a minimum of two peak time interval is required.
|
|
* Just note the timestamp of this peak.
|
|
*/
|
|
if(firstTime == true)
|
|
{
|
|
firstTime = false;
|
|
previousBeatTime = WatchdogTimer_GetTimestamp();
|
|
}
|
|
else
|
|
{
|
|
/* Rolling window of two samples. Note the timestamp of
|
|
* the new peak and subtract the timestamp of the previous
|
|
* to obtain the RR-interval. Extrapolate it to get a heart
|
|
* beat value in beats per minute.
|
|
*/
|
|
newBeatTime = WatchdogTimer_GetTimestamp();
|
|
twoSampleTime = newBeatTime - previousBeatTime;
|
|
|
|
if(twoSampleTime != 0)
|
|
{
|
|
heartRate = (uint32)SEC_IN_MIN * MS_TO_SECOND / twoSampleTime;
|
|
}
|
|
|
|
previousBeatTime = newBeatTime;
|
|
}
|
|
}
|
|
|
|
/* Clear the flag to indicate next time that this R peak has already
|
|
* been accounted for and we need to wait for the next peak.
|
|
*/
|
|
newBeat = false;
|
|
}
|
|
else
|
|
{
|
|
/* Set the flag to indicate that there is no R-peak going on right now.
|
|
* So it is expected that the next time the ADC output is more than
|
|
* the threshold, it will be a new peak.
|
|
*/
|
|
newBeat = true;
|
|
}
|
|
}
|
|
|
|
|
|
/* [] END OF FILE */
|